Using ETC (Every Test Calibration) real-time calibration technique, in each measurement, the spectrophotometer will be to calibrate the system hardware and software. This technique solves the problem of long-term stability and measurement data drift color measuring instrument fundamentally. Spectral reflectance standard deviation within 0.08, Chroma value Delta E*ab maximum value of 0.03, the Inter-instrument agreement is less than 0.2.
Illumination | d/8(Diffused lighting, 8 degrees observe angle)(conform to CIE No.15、ISO 7724/1、ASTM E1164、DIN 5033 Teil7、JIS Z8722 Condition c standards) | |
Size of integrating sphere | Φ40mm,Alvan diffused reflection surface coating | |
Illumination Light source | CLEDs(entire wavelength balanced LED light source) | |
Sensor | dual light path sensor array | |
Wavelength range | 400-700nm | |
Wavelength interval | 10nm | |
Half spectral width | 5nm | |
Reflectivity range | 0-200% | |
Reflectivity resolution | 0.01% | |
Observation angle | 2°/10° | |
Measurement light source | A,C,D50,D55,D65,D75,F1,F2,F3,F4,F5,F6,F7,F8,F9,F10, F11,F12,DLF,TL83,TL84,NBF,U30,CWF |
|
Data being displayed | SPD distribution/data, sample's color values, color difference values/graph, pass/fail results, color error tendency, color simulation, display measurement area,history data color simulation, manual input standard sample, generate measurement report | |
Measurement time | About 1 second | |
Measurement time interval | 0.5 second | |
Measuring caliber | Φ11mm | |
Color space | CIE-L*a*b,L*C*h,L*u*v,XYZ,Yxy,reflectivity,Hunterlab,Munsell MI,CMYK | |
Color difference formulas | ΔE*ab,ΔE*CH,ΔE*uv,ΔE*cmc(2:1),ΔE*cmc(1:1),ΔE*94,ΔE*00,ΔEab(Hunter),555,color classification | |
Other colorimetric indices | WI(ASTM E313-00,ASTM E313-73,CIE/ISO,AATCC,Hunter,Taube Berger Stensby),YI(ASTM D1925,ASTM E313-00,ASTM E313-73), Tint(ASTM E313,CIE,Ganz),metamerism index Milm, adhesive/changing color fastness, ISO luminance,8 gloss, A density, T density,E density, M density |
|
Repeatability | light splitting reflectivity:standard deviation within 0.08% | |
color values:ΔE*ab<=0.015(After calibration, standard deviation of 30 measurements on test white board, 5 second intervals),Maximum:0.03 | ||
Battery capacity | rechargeable, 20000 continuous tests, 7.4V/6000mAh | |
Interface | USB | |
Data storage | 40000 test results | |
Light source longevity | 10 years, 3 million tests | |
Inter-instrument agreement | ΔE*ab within 0.2(BCRA color charts II, average of the 12 charts) | |
Size | 350*300*200mm(L*W*H) | |
Weight | About 4.5kg | |
Display | 5-inch True color screen | |
Work temperature range | 0~45℃, relative humidity 80% or below(at 35°C),no condensation | |
Storage temperature range | -25℃to 55℃,relative humidity 80% or below(at 35°C),no condensation | |
Standard accessories | AC power line, manual, color management software, drive software, electronic manual, color management guide, USB cable, black/white calibration tube, electronic color charts | |
Optional accessories | powder molding device, measurement and test report | |
Color matching system | Matches | |
UV light source | Does not include UV light source |